Aeroqual Volatile Organic Compound (VOC) Modules

PID gas analyzer module with integrated zero cycle and nafion membrane
Dual range – 0 – 30 ppm, 0 – 500 ppb

$4,050.00

Availability: 8 in stock

Product Information

Volatile organic compounds are a group of compounds that can occur as gases in ambient air. Aeroqual’s VOC modules provide continuous measurements for perimeter monitoring, with low- and high-range options to suit different expected concentrations.

Volatile Organic Compounds (VOCs) is the name given to a group of organic compounds which can exist as a gas in the atmosphere.

VOCs are not listed among the six criteria health pollutants cited by the U.S. EPA. However, the effects of VOC exposure are becoming more understood and demands for ambient VOC monitoring are increasing.

Product Features

Specification Low-range VOC High-range VOC
Measurement technology Photoionization detector – PID Photoionization detector – PID
Measurement range 0-500 ppb 0-30 ppm
Display resolution 0.1 ppb 0.01 ppm
Lower detection limit <1 ppb <0.1 ppm
Precision 2% of reading or 1 ppb 2% of reading or 0.05 ppm
Noise: Zero; Span (% of reading) <1 ppb; 1% <0.1 ppm; 1%
Linearity (% of FS) 1% 2%
Drift 24 hour: Zero; Span (% of FS) 1 ppb; 1% 0.1 ppm; 1%
Monitor platforms AQS 1 and AQM 65 AQS 1 and AQM 65

These are alternative module ranges, not a single selectable-range specification.

Performance & Field Data

Aeroqual reports R² = 0.99 in controlled multipoint gas tests for both ranges. Measurements are typically calibrated to isobutylene. Response varies by compound: the module does not identify individual VOCs, provide benzene-selective results or detect methane.

Results shown are from the specific manufacturer-reported evaluation described. Performance depends on the monitoring configuration, calibration, site conditions and measurement objectives.

Typical Applications

  • Remediation perimeter monitoring
  • Investigations around fuel- and solvent-related activities
  • Industrial VOC event assessment
  • Select the range and calibration approach for the anticipated mixture and project objectives.

Select the right module for your project

Tell Specto what you need to measure, the concentrations you expect and the monitor you plan to use. Our engineers and scientists can help you select the module, configuration and calibration approach for your project.

Discuss Your Monitoring Requirements

Datasheet(s)

Aeroqual AQS1 Datasheet

Manuals

Aeroqual AQS DS Module Technical Performance Guide