Noise Academy
Sound Level Meter Calibration
process that establishes confidence in the relationship between the sound
reaching a microphone and the result reported by the measurement system.
ON THIS PAGE
In practice, the word calibration is used for several different activities: applying an acoustic calibrator in the field, adjusting an instrument, periodically testing it in a laboratory, and evaluating whether an instrument design meets a standard.
Those activities are related, but they are not interchangeable.
For a short attended survey, measurement quality may be demonstrated with an acoustic check before measurement and another afterward. For a noise monitor operating continuously for months or years, that model is not enough. Long-term monitoring requires an ongoing chain of system checks, external verification, preventive service and periodic laboratory testing.
Understanding those differences is particularly important when measurements may be reviewed by a client, regulator or independent technical expert.
Calibration, Checks, Adjustment and Periodic Testing
The terminology matters because each activity answers a different question.
| Activity | What it tells you | Typical application |
|---|---|---|
| Field acoustic check | Compares the complete measurement chain with a known acoustic reference | Before/after surveys and periodically during long-term monitoring |
| Adjustment | Changes the instrument's sensitivity or calibration factor to agree with the reference | When required before measurement or following service |
| Automated system check | Verifies aspects of the measurement path without necessarily changing the calibration factor | Unattended monitoring |
| Factory calibration / production verification | Documents the condition of an individual serial-numbered instrument when supplied or serviced | New instruments and manufacturer service |
| IEC 61672-3 periodic testing | Tests an individual sound level meter against a defined set of key performance requirements | Periodic laboratory verification |
| Pattern or type evaluation | Evaluates an instrument design against the applicable performance requirements | Model-level approval |
Calibration Is Not the Same as Adjustment
In metrology, calibration determines the relationship between an indication and a reference.
Adjustment changes the instrument.
A field acoustic check may therefore show that an instrument agrees with its reference, or it may indicate that an adjustment is needed.
For measurements where drift is being evaluated, this distinction becomes especially important. An as-found result should be recorded before changing the calibration factor, otherwise useful evidence about the preceding measurement period can be lost.
What Does Class 1 Mean?
A sound level meter is designed to meet either Class 1 or Class 2 performance specifications under IEC 61672-1.
Independent pattern or type evaluation under IEC 61672-2 provides evidence that a particular instrument design meets those specifications.
That approval applies to the instrument model or design. It should not be confused with the periodic calibration status of an individual serial-numbered meter.
The SVANTEK SV 307A, for example, is a Class 1 noise monitoring station under IEC 61672-1:2013, and current SVANTEK information identifies the model as independently type approved.
An individual SV 307A does not cease to be a Class 1 model simply because an optional accredited IEC 61672-3 certificate was not purchased with that particular unit.
The distinction is:
- Class 1 describes instrument performance/design
- Periodic accredited testing provides additional evidence about the condition of a particular instrument at a particular time.
Factory Certificate vs Accredited IEC 61672-3 Testing
SVANTEK provides two different levels of documentation that are useful to distinguish.
Factory Calibration Certificate
A factory calibration certificate supplied with an instrument is a serial-number-specific production and calibration record.
According to information provided directly to Specto by SVANTEK, its factory certificate documents production verification under the manufacturer’s quality system and provides traceability to the applicable national reference standard.
It confirms the condition of that individual instrument when tested by the manufacturer.
It is not the same document as an accredited IEC 61672-3 periodic-test certificate.
Accredited IEC 61672-3 Periodic Testing
IEC 61672-3 defines procedures for periodic testing of individual sound level meters.
The purpose is to provide confidence that the instrument meets the applicable IEC 61672-1 requirements for a limited set of key tests under the conditions of the periodic evaluation. IEC specifically notes that Part 3 is not a repeat of the complete pattern-evaluation process.
An accredited periodic-test certificate can add:
- serial-number-specific test results
- stated measurement uncertainty
- accredited laboratory status
- metrological traceability
- independent evidence that the instrument met the applicable periodic-test tolerances when tested
This level of documentation can be particularly valuable where it is required by a client, permit, standard or quality plan, or where the measurements are expected to receive a higher level of scrutiny.
Acoustic Calibrators and IEC 60942
A field acoustic check uses a device that generates a known sound pressure level at the microphone.
These sound calibrators are covered by IEC 60942.
IEC 60942:2017 defines Laboratory Standard, Class 1 and Class 2 calibrators. A Class 1 sound calibrator is primarily intended for use with a Class 1 sound level meter.
Most field calibrators use a nominal frequency of 1 kHz.
Common reference levels are:
94 dB and 114 dB
A 94 dB sound pressure level corresponds approximately to 1 Pa relative to the standard reference pressure of 20 µPa.
The value used for the check should follow the calibrator’s current certificate and the manufacturer’s instructions rather than assuming that every calibrator produces exactly its nominal value under every condition.
Example: SVANTEK SV 36
The SV 36 is a Class 1 acoustic calibrator under IEC 60942:2017 and produces 94 dB or 114 dB at 1 kHz.
It uses internal reference, atmospheric-pressure and temperature sensing to control the generated sound level automatically.
That is important because environmental correction depends on calibrator design. A generic rule such as “apply an altitude correction above 3,000 feet” should not be applied to every modern calibrator.
Always follow the instructions for the actual calibrator being used.
Calibration Checks for Short Attended Measurements
For an attended or relatively short measurement, the traditional workflow is straightforward:
- Initial acoustic check
- measurement
- final acoustic check without adjustment
Before measurement, the sound calibrator is fitted correctly to the microphone and the complete measurement channel is checked.
If adjustment is permitted or required by the applicable procedure, it should occur before the measurements used for assessment.
At the end of the measurement series, the calibrator is applied again.
This final check should normally be recorded without first adjusting the instrument, because its purpose is to determine whether the measurement chain changed during the survey.
The difference between the initial and final results provides evidence about measurement stability.
The 0.5 dB Criterion Is Method-Specific
A 0.5 dB difference is commonly encountered in standards, but it should not be presented as a universal calibration rule.
For example, ISO 1996-2:2017 specifies that the difference between consecutive calibration checks should be no greater than 0.5 dB for its environmental-noise measurement procedure. The standard remains current following review in 2026.
Other measurement methods may impose different procedures or acceptance criteria.
The applicable standard, permit or project specification should therefore control the decision.
Long-Term Unattended Monitoring Requires a Different Approach
A monitoring station deployed for six months, one year or five years should not be treated as one enormous measurement with a calibration check only at installation and final removal.
If an instrument passed its check in 2027 and failed its next check in 2032, there would be little evidence showing when the measurement chain changed.
Instead, long-term monitoring should be supported by a continuing sequence of verification events.
A practical long-term workflow is:
- Installation reference
- automated system checks
- periodic external field verification
- preventive service
- periodic laboratory testing
- final as-found check
Each successful verification point helps establish confidence in the measurement period since the preceding check.
That effectively divides a multi-year project into traceable measurement intervals.
If a Check Shows Significant Change
Suppose a monitoring station passes an external check in January, passes again in July, but shows an unacceptable difference during the next scheduled visit.
The correct workflow is:
- Record as-found result
- evaluate preceding measurement interval
- investigate
- adjust or service if required
- establish new reference
- resume measurement
The instrument should not simply be adjusted immediately and returned to service without preserving its as-found condition.
That record is what allows the previous measurement period to be evaluated separately from the period that follows.
Automated System Checks in Unattended Monitoring
Modern unattended noise monitors can provide another layer of quality assurance between technician visits.
The SV 307A, for example, incorporates a remote system-check function and a built-in acoustic source. Current product documentation describes a patented system-check arrangement designed to support ongoing monitoring of the measurement path.
SVANTEK’s operating documentation distinguishes a system check from calibration: the check can assess the measuring path without changing the instrument’s calibration factor. The manufacturer also recommends a 12-month factory calibration interval to maintain confidence in continuing accuracy.
This distinction is important.
An automated system check provides ongoing evidence of instrument health.
It does not replace every external acoustic verification or laboratory periodic test.
Instead, the different controls work together:
automatic checks detect problems quickly
external checks provide an independent acoustic reference in the field
laboratory testing evaluates additional instrument performance
Specto's Recommended Approach for Long-Term Monitoring
Different projects have different regulatory, contractual and quality requirements, so there is no single maintenance interval appropriate to every installation.
For professional unattended environmental monitoring, Specto’s general recommended approach is:
| Frequency region | Approximate range |
|---|---|
| Automated system checks | Use regularly where supported by the monitoring system and appropriate to the project |
| Scheduled external field verification and preventive service | Approximately every 3–6 months for long-duration unattended installations |
| Sound calibrator laboratory calibration | Annually as general best practice |
| Accredited sound level meter periodic testing | Annual preferred |
| Extended laboratory interval | Up to 2 years where governing requirements permit and regular documented verification demonstrates stable performance |
| Where regular external field verification is not performed | Specto recommends annual laboratory testing |
The 3–6 month visit is not simply a calibration appointment.
For an unattended station it can also include inspection of:
- microphone and environmental protection
- mounting and physical security
- external power or solar equipment
- local storage
- overall station condition.
- windscreen condition
- cables and connectors
- communications
- system-check history
Calling this activity scheduled field verification and preventive service more accurately describes what long-term monitoring requires.
Why One or Two Years?
ISO 1996-2 recommends testing measurement-system performance once per year and identifies two years as the maximum interval between those laboratory verifications for its environmental-noise framework.
Specto therefore treats annual testing as the preferred baseline.
Where the applicable requirements permit it, extending an instrument to a two-year laboratory interval can be reasonable when there is:
- a stable calibration and service history
- regular documented external acoustic verification
- ongoing automated system checks where available
- no evidence suggesting that the measurement chain has changed.
A two-year interval should not be described as something automatically “earned” merely by performing field calibration.
Field checks and IEC 61672-3 laboratory testing examine different aspects of measurement assurance.
If regular external field verification is not taking place, Specto recommends retaining the annual laboratory interval.
Project specifications, regulatory requirements and client quality plans always take precedence over these general recommendations.
Periodic Testing Does Not Replace Type Approval
The IEC 61672 series separates instrument design from periodic testing.
IEC 61672-1
defines sound level meter performance specifications.
IEC 61672-2
addresses pattern-evaluation tests used to assess instrument designs.
IEC 61672-3
addresses periodic testing of individual instruments.
IEC 61672-3 intentionally evaluates only a limited set of key characteristics. Passing those periodic tests is valuable evidence about the individual unit, but it is not equivalent to repeating the full model-level pattern evaluation.
This is why an independently type-approved Class 1 instrument with documented periodic verification provides two different pieces of evidence:
the design has been independently evaluated, and the individual instrument has subsequently been checked.
Laboratory Accreditation and Traceability
Laboratory certificates should identify the instrument, the procedure used, the measured results and the measurement uncertainty where applicable.
For work requiring accredited calibration, the laboratory should be accredited to ISO/IEC 17025 for the relevant tests and measurements.
In the United States, calibration results are often described as traceable to NIST.
Traceability does not necessarily mean that NIST directly calibrated the customer’s sound level meter.
Instead, metrological traceability is established through a documented, unbroken chain of calibrations to an appropriate reference, with each step contributing to the stated measurement uncertainty.
The accreditation scope matters as much as the accreditation name.
A laboratory may be ISO/IEC 17025 accredited without necessarily being accredited for the specific acoustic measurement or instrument being submitted.
The Records That Support Long-Term Data
A good calibration and verification record should make it possible for someone reviewing the project later to reconstruct the measurement-quality history.
For a long-term unattended deployment, useful records include:
- instrument and microphone identification and serial numbers
- relevant factory and accredited laboratory certificates
- acoustic-calibrator identification and certificate
- installation/reference check
- as-found results from scheduled field visits
- any adjustments made and the new reference established afterward
- automated system-check history
- service and component-replacement records
- periods when instrument condition was uncertain
- decisions to exclude or retain affected measurement data
- the reason for those decisions
This turns calibration from an isolated certificate into a traceable measurement history.
For a five-year monitoring program, that history is substantially more useful than a certificate from year one and another result taken when the equipment is finally removed.
The Practical Takeaway
Calibration quality is not defined by how often someone presses a calibration button.
It comes from understanding what each part of the process establishes.
For a short attended survey:
- check
- measure
- final check
may be sufficient when it follows the applicable measurement procedure.
For long-term unattended monitoring:
- reference
- monitor
- verify
- service
- periodically test
- document
is the more appropriate model.
Automated checks, external field verification, factory documentation and accredited IEC 61672-3 periodic testing each provide different evidence. None should automatically be treated as a substitute for all the others.
The objective is to maintain a continuous, documented chain of confidence in the measurement system for as long as the monitoring project operates.
For related topics, see Noise Monitoring, Class 1 vs Class 2 Sound Level Meters, Occupational Noise Exposure, Sound Measurements, and the Specto Noise Academy.