Noise Academy
Class 1 vs Class 2
Sound Level Meters
Why instrument class — and independent pattern approval — matter when
noise data has to stand up to scrutiny
ON THIS PAGE
Class 1 and Class 2 are the two performance categories defined for sound level meters under IEC 61672-1. Both are legitimate measurement-instrument classes. The principal difference is that Class 1 instruments must meet tighter acceptance limits and a more demanding environmental performance specification than Class 2 instruments.
That does not mean Class 2 instruments are inherently unsuitable or inaccurate. They can be entirely appropriate for many occupational, screening and general-purpose measurements.
The more important question for many environmental noise projects is:
How much evidence do you need that the instrument actually performs to the claimed class?
That is where pattern approval, also commonly called type approval, becomes important.
A manufacturer may state that an instrument conforms to the Class 1 specifications of IEC 61672-1. A pattern-approved Class 1 meter goes further: the instrument design has undergone the independent testing defined by IEC 61672-2 to verify its performance against the applicable Part 1 requirements.
For monitoring associated with permits, environmental assessments, community complaints, major infrastructure projects or other work where results may be closely examined, that independent verification can provide an additional level of confidence.
Class 1 and Class 2: What Does the Standard Actually Say?
IEC 61672-1 defines two performance categories for sound level meters: Class 1 and Class 2.
They share the same basic measurement objectives. The difference is primarily how tightly the instrument must remain within the standard’s permitted limits.
In practical terms:
| Class 1 | Class 2 | |
|---|---|---|
| Performance limits | Tighter | Wider |
| Environmental performance requirements | More demanding | Less demanding |
| Typical role | Precision and professional measurements | General-purpose measurements |
| Suitable where specified by governing method | Yes | Yes, where permitted |
| Pattern approval available | Yes | Can be, depending on instrument/model |
It is therefore better to think of Class 1 and Class 2 as different levels of metrological performance, rather than “good” and “bad.”
The right class should always begin with the governing specification, regulation, standard or monitoring plan.
A Class 1 Claim and a Pattern-Approved Class 1 Instrument Are Not the Same Thing
This is one of the most important distinctions when evaluating a sound level meter.
IEC 61672 is divided into three related parts:
IEC / ANSI Part 1
What performance must the instrument achieve?
IEC / ANSI Part 2
Has the instrument design been independently evaluated?
IEC / ANSI Part 3
Does this individual instrument continue to perform correctly?
| Standard | Purpose | What it tells you |
|---|---|---|
| IEC 61672-1 | Specifications | Defines the performance requirements for Class 1 and Class 2 sound level meters |
| IEC 61672-2 | Pattern evaluation tests | Defines the tests used to evaluate whether an instrument design meets the applicable Part 1 requirements |
| IEC 61672-3 | Periodic tests | Provides a reduced test set for evaluating individual instruments periodically |
A manufacturer’s Class 1 conformance statement tells you the instrument is specified to meet the Class 1 requirements.
Pattern approval provides independent evidence that the instrument design has actually undergone the much broader Part 2 evaluation process.
That distinction matters because Part 2 evaluation goes well beyond checking a meter at one sound level or conducting a routine acoustic calibration. It evaluates the instrument as a system across the performance and influence factors addressed by the standard.
Why Pattern Approval Matters
A sound level meter does not operate in an ideal laboratory once it reaches a project.
Its microphone, electronics, housing, accessories and configuration can all influence its response. Environmental and electromagnetic conditions can also affect measurement performance.
Pattern evaluation is designed to test the instrument model, rather than simply checking one individual meter at a calibration point.
For the end user, this provides three important advantages.
01
Independent verification.
The manufacturer’s performance claim has been evaluated independently against a defined test standard.
02
Confidence in the complete instrument configuration.
Pattern approval applies to the approved model and specified configuration, not merely an isolated microphone or electronic component.
03
Stronger measurement documentation.
Where a monitoring program is reviewed by a client, agency, consultant or other third party, an approval certificate provides additional evidence supporting the selection of the instrument.
Pattern approval should not, however, be treated as a guarantee of regulatory compliance or legal admissibility.
Instrument selection is only one part of producing defensible noise data. Microphone placement, field calibration, meteorological conditions, monitoring procedures, instrument settings, maintenance, data review and compliance with the applicable monitoring method all remain important.
Does Pattern Approval Matter in the United States?
The distinction needs some context in North America.
The United States does not have one centralized federal pattern-approval requirement for sound level meters comparable with the legal-metrology systems used in several European countries. Svantek describes Part 2 pattern approval as voluntary in the U.S. rather than a general federal requirement.
However, the technical standards themselves are closely aligned.
ANSI/ASA S1.4-2014 adopted IEC 61672 in the United States, including separate standards covering specifications, pattern evaluation and periodic testing.
That means an independent approval issued against IEC 61672 can still be valuable evidence for a U.S. buyer evaluating whether a manufacturer’s Class 1 claim has been independently tested.
The important distinction is:
Pattern approval may not be universally required in the United States, but it provides a higher level of independent verification than a specification claim alone.
Always check the actual project specification, agency requirement and monitoring method before determining what approval or instrument class is required.
When Is Class 2 Appropriate?
There are many situations where Class 2 can be a sensible choice.
For example, OSHA’s technical guidance describes legacy ANSI Type 2 sound level meters as general-purpose instruments and states that Type 2 meters provide the minimum precision OSHA requires for many workplace noise measurements.
Class 2 or equivalent general-purpose instruments may therefore be appropriate for applications such as:
- preliminary noise surveys and screening;
- identifying areas requiring additional investigation;
- internal facility assessments;
- comparative measurements;
- occupational noise applications where the applicable method allows them;
- training and education; and
- projects where the governing specification explicitly permits Class 2 performance.
The governing method always takes priority.
A Class 2 instrument used correctly for an appropriate application can provide much more useful information than a higher-specification instrument used incorrectly.
When Should You Favor Class 1?
Class 1 becomes increasingly important as the measurement uncertainty, environmental conditions or consequences of a disputed measurement become more significant.
Typical examples include:
- environmental and community noise monitoring;
- construction noise monitoring where Class 1 is specified;
- environmental impact and baseline studies;
- permit or contract monitoring requiring Class 1 performance;
- measurements associated with community complaints;
- long-term unattended monitoring;
- detailed acoustic investigations; and
- projects where measurement results may receive substantial technical or regulatory scrutiny.
The tighter performance requirements of Class 1 give the measurement team more confidence that the instrument itself is not introducing unnecessary uncertainty.
And where the project justifies it, selecting a pattern-approved Class 1 instrument provides additional independent evidence behind that performance claim.
Class 1 vs Type 1: A Terminology Note
“Type 1” still appears frequently in U.S. specifications.
It comes from earlier versions of ANSI S1.4, which used Type 1 and Type 2 terminology. The current ANSI/ASA standard is based on IEC 61672 and uses Class 1 and Class 2 instead. ANSI identifies its current Part 1 standard as the successor to ANSI S1.4-1983.
The terms are often treated as broadly comparable in project specifications, but they are not simply interchangeable editions of the same requirement.
If a contract specifically calls for “ANSI Type 1,” it is worth confirming which standard edition the specification intends rather than assuming.
What Does This Mean for Outdoor Noise Monitoring?
For professional environmental monitoring, instrument class should be considered as part of the complete monitoring system.
Ask:
- What class does the project specification require?
- Is the manufacturer's Class 1 claim supported by independent pattern/type approval?
- Does the approval cover the exact instrument model and required configuration?
- What calibration and periodic-testing documentation is available?
- Is the instrument designed for long-term outdoor deployment?
- Can the system remotely report status, measurement data and diagnostic information?
- Are the required frequency analysis, audio, reporting and alert capabilities supported?
These questions tell you much more than a Class 1 logo on a datasheet.
The SVANTEK SV 307A: Class 1 With Independent Type Approval
The SVANTEK SV 307A is an example of the distinction.
Svantek’s current specifications identify the SV 307A as a Class 1 sound level meter under IEC 61672-1:2013 with PTB Type Approval. Svantek also identifies it as the first Class 1 noise monitoring station using its MEMS microphone technology to obtain PTB type approval.
For projects requiring long-term environmental noise monitoring, the value is therefore not simply that the instrument carries a Class 1 specification.
Its Class 1 performance has also been subjected to an independent pattern-approval process.
That gives project teams an additional layer of technical evidence when they need to explain why a particular instrument was selected.
Choosing the Right Sound Level Meter
The objective should not be to specify Class 1 for every measurement.
It should be to select an instrument whose performance and supporting evidence are appropriate for the decisions being made.
For general-purpose measurements, a suitable Class 2 instrument may provide everything the project requires.
For professional environmental monitoring, projects governed by Class 1 specifications, or measurements that may face significant external scrutiny, there is a strong case for going further:
Specify Class 1 — and consider whether that Class 1 claim is backed by independent pattern approval.
Because when the data matters, understanding how the instrument’s performance was verified can be just as important as the class printed on its datasheet.
For related topics, see Sound Level Meter Calibration, Background Noise and Residual Sound, Noise Monitoring, and the Specto Noise Academy.