Product Information
Nitrogen dioxide is associated with combustion and can also form through atmospheric reactions. The Aeroqual NO₂ module provides continuous ambient measurements for projects assessing traffic-related pollution, industrial impacts and local air quality.
Product Features
| Specification | Details |
| Measurement technology | Electrochemical with integrated O3 filter and ABC |
| Measurement range | 0-500 ppb |
| Display resolution | 0.1 ppb |
| Lower detection limit | <1 ppb |
| Precision | 2% of reading or 2 ppb |
| Linearity | 1.5% of full scale |
| Noise (zero / span) | < 1 ppb / 1% of reading |
| 24-hour drift (zero / span) | 1 ppb / 0.2% of full scale |
| Monitor platforms | AQS 1 and AQM 65 |
The integrated ozone filter and sample/zero cycling help manage interference and baseline drift.
Performance & Field Data
- Co-located with a regulatory station in Riverside, CA (April 2022): R² = 0.96, mean absolute error = 1.3 ppb.
- EU Guide to Demonstration of Equivalence total measurement uncertainty of 5.8% at a 100 ppb hourly limit value, within the 15% regulatory-grade threshold (Directive 2008/50/EC).
- Recommended field calibration interval of 1–3 months.
Results shown are from the specific manufacturer-reported evaluation described. Performance depends on the monitoring configuration, calibration, site conditions and measurement objectives.
Typical Applications
- Urban and national ambient air quality monitoring
- Roadside and near roadside monitoring
- Coal-fired power station perimeter monitoring
- Mine perimeter monitoring
- Oil and gas perimeter monitoring
- Checking pollution ‘hotspots’
- Air quality research
Select the right module for your project
Tell Specto what you need to measure, the concentrations you expect and the monitor you plan to use. Our engineers and scientists can help you select the module, configuration and calibration approach for your project.