Product Information
Hydrogen sulfide is commonly associated with wastewater treatment and waste handling. The Aeroqual H₂S module measures low-level ambient concentrations, helping teams investigate H₂S events and changes around site boundaries.
Product Features
| Specification | Details |
| Measurement technology | Electrochemical sensor with automatic baseline correction |
| Measurement range | 0-5,000 ppb – equivalent to 0-5 ppm |
| Display resolution | 0.1 ppb |
| Lower detection limit | 2 ppb |
| Precision | 1% of reading or 3 ppb |
| Linearity | 0.5% of full scale |
| Noise (zero / span) | 1 ppb / 0.1% of reading |
| 24-hour drift (zero / span) | < 1 ppb / < 0.5% of full scale |
| Monitor platforms | AQS 1 and AQM 65 |
Sample/zero cycling helps reduce baseline drift during ambient monitoring.
Performance & Field Data
Aeroqual reports R² = 0.99 in a controlled six-point gas test spanning 0-1,600 ppb. This demonstrates response linearity under the test conditions; it is laboratory validation, not an outdoor co-location result.
Results shown are from the specific manufacturer-reported evaluation described. Performance depends on the monitoring configuration, calibration, site conditions and measurement objectives.
Typical Applications
- Wastewater treatment plant boundaries
- Waste handling and disposal sites
- Investigations of H₂S-related concentration events
Select the right module for your project
Tell Specto what you need to measure, the concentrations you expect and the monitor you plan to use. Our engineers and scientists can help you select the module, configuration and calibration approach for your project.